9. Disconnect the anode signal wire from the
detector.
Detector top
cover
Remove anode signal wire
Figure 1-10. Removing the cover, anode purge fitting, vent tube, and anode signal wire
10. Fully loosen the four Torx T-20 screws on
the detector pallet and lift the pallet and
detector from the GC.
Released: April 2004
11. Thread the heater/sensor leads out of the
6850 Series Gas Chromatograph Service Procedures
detector pallet and remove the detector from
the pallet.
Anode purge line
Makeup gas line
Remove µECD vent
tube