Deactivated Splitless Liner
Part number 5062-3587
900 µL internal volume to
expansion accommodate solvent
Glass wool at base to filter non-volatile
sample components
Larger outside diameter to limit sample
exposure to injection port body during
the splitless delay.
Released: April 2004
6850 Series Gas Chromatograph Service Procedures
Deactivated Split Liner
Part number 5183-4647
Glass wool held in place by indents to
provide complete sample vaporization and
to wipe the syringe needle.
Glass bead at the base to provide optimum
clearance for split flow with minimal
restriction.
Smaller outside diameter than splitless
liners.