Download Print this page

Agilent Technologies 6850 Series Service Information page 263

Gc system

Advertisement

1. Select Status / Setup / Configure, then
press Edit.
2. Select Edit Oven Type.
Released: April 2004
3. Enter the password as follows: Press dot (.)
4. Select the correct oven type, then press OK.
6850 Series Gas Chromatograph Service Procedures
Caution
Be sure the GC is physically configured for
the oven type you select. A 200–240 V
instrument will perform abnormally if set to
operate as a 100–120 V oven, while a 100–
120 V oven can be dangerously overheated if
set to operate as a 200–240 V oven.
then left arrow ( ) to erase the question
mark (?), then enter ". . ." (three dots)
as the password. Once this is done, the set-
tings options are enabled.