To remove the oven intake fan:
1. Turn off the instrument and remove the
power cord.
2. Remove the front cover. See
(with keyboard and display)"
WARNING
Do not unbalance the GC. It is heavy and
might fall on your head.
3. Slide the front of the GC just far enough over
the edge of the benchtop to gain access to
the intake fan.
Released: April 2004
4. Remove three of the four T-20 Torx screws
"Front cover
on page 1-20.
5. While supporting the fan with one hand,
6. Support the fan with one hand and remove
6850 Series Gas Chromatograph Service Procedures
that hold the fan in place.
remove the last screw.
the grill assembly. Note the orientation of
the grill inlets—they are offset so that the
grill extends down.