5. Remove the gang block fitting from the inlet
flow module.
If a valve box is installed, you will need to
remove its top cover to access the gang
fitting.
Inlet flow
module
Gang block
fitting
Released: April 2004
6. Remove the split vent trap bracket.
7. Disassemble the split vent trap.
8. Remove the inlet insert weldment and
6850 Series Gas Chromatograph Service Procedures
attached parts from the instrument.
Split vent trap
Split vent trap
bracket
62