4. Screw the column nut on fingertight. Use a
wrench to tighten the column nut 1/8 to
1/4 turn. Be careful not to overtighten.
Removing the septumless head
1. Cool the inlet to room temperature.
2. Disconnect the carrier gas line.
3. Unscrew the septumless head counterclock-
wise from the inlet.
4. Screw the new head onto the inlet. Tighten
it 1/8 turn past finger tight.
Released: April 2004
Figure 1-39. Removing the carrier gas line
5. Reconnect the carrier gas line.
6850 Series Gas Chromatograph Service Procedures
108