Agilent Technologies 6850 Series Service Information page 113

Gc system

Advertisement

3. Remove the septum or Merlin microseal,
taking care not to scratch the interior of the
septum head.
4. Install a new septum or Merlin microseal and
the correct cap. When installing a Merlin
microseal, note that the side with the visible
metal parts faces the oven.
Figure 1-42. PTV septa
5. Check for leaks out of the cap and tighten
the cap if necessary.
Replacing the liner
1. Remove the head from the inlet.
Released: April 2004
2. Grasp the liner by the Graphpack ferrule.
3. Unscrew the assembly tool (part no. 5182-
Figure 1-43. Installing the ferrule onto the
6850 Series Gas Chromatograph Service Procedures
Remove the liner and ferrule.
9750) the ferrule guide from the compression
fitting.
new liner
113