Agilent Technologies 6850 Series Service Information page 170

Gc system

Advertisement

• The end of the column does not protrude
from the top of the makeup gas adapter.
For most columns (outer diameter
> 0.15 mm), insert the column as far as
it will go into the gigabore liner. If using
a microbore column that passes com-
pletely through the liner, position the
column so that the total length from the
back of the capillary nut to the end of
the column is about 70 to 72 mm.
Detector interface card
The µECD detector has an interface board off of
the main board as shown below.
Caution
Make sure you are properly grounded with an
ESD strap before continuing.
Released: April 2004
1. Unlock and unclip the ribbon cable from the
2. Remove the screw at the top of the card
6850 Series Gas Chromatograph Service Procedures
interface board.
where it mounts to the main board and pull
the card out.