Agilent Technologies 6850 Series Service Information page 261

Gc system

Advertisement

Caution
When you update the GC firmware, all
method and LAN information is erased. To
avoid this, copy/record methods and LAN
information, update the GC firmware, then
reload the saved material.
3. Select the update file and press Execute.
To update the Injector firmware, the Control
Module must contain firmware revision
A.03.00 or higher.
Released: April 2004
4. When the confirmation screen appears, reply
5. For GC firmware updates, restore the meth-
6850 Series Gas Chromatograph Service Procedures
Caution
Do not turn off the GC power, disconnect the
GC power cord, or disconnect the Control
Module from the GC until the update process
is either completed or cancelled. A disruption
to the Control Module during the firmware
update can make the GC unusable.
Yes to continue or No to cancel this proce-
dure.
The firmware will be updated. The update
may take several minutes. When complete,
the instrument restarts to enable the new
firmware.
ods and local LAN addressing information (if
used).
262