Replacement Procedures
Injection port cooling fan
WARNING
Shock hazard. Before beginning, turn off the
instrument and disconnect the power cord.
1. Turn off the instrument and remove the
power cord.
2. Remove the top cover. See
page 1-23.
3. Unscrew three T-10 Torx screws and remove
the fan.
Released: April 2004
4. Disconnect the fan wires at the connector.
When reinstalling the fan, make sure to install
the fan so that the fan exhaust port faces the
inlet.
"Lid top cover"
on
6850 Series Gas Chromatograph Service Procedures
Fan wire
connector
Port on side
(not visible)