Agilent Technologies 6850 Series Service Information page 192

Gc system

Advertisement

8. Use two wrenches to loosen the Swagelok
connection between the jet assembly and the
heated transfer tube.
9. Carefully lift the detector module and jet
assembly off of the transfer tube and out of
the GC. Be careful not to damage the fused
silica gigabore liner.
It is not necessary to disconnect any plumb-
ing or ignitor leads. Leave all attached and
disconnect the detector block from the trans-
fer line at the Swagelok fitting, then gently
lift the block and set aside.
10. Remove the heater/sensor assembly and
install a new one.
11. Reassemble in reverse order.
Released: April 2004
Ignitor glow plug
1. Turn off power to the gas chromatograph
2. Allow time for heated zones to cool to safe
3. Pull the PMT housing off the detector
6850 Series Gas Chromatograph Service Procedures
and disconnect the main power cord. Remove
the detector covers.
temperatures.
Caution
Always turn the electrometer or the main
power off before removing the PMT housing
to avoid destroying the tube.
module and remove the filter from the detec-
tor. Set both in a safe place. Keep the open
end of the PMT covered whenever practical
to avoid light damage to the tube.
192