To install the main board
When installing the main board, refer to the
following figure for the board's pinouts.
If installing a new main board, configure it for
the correct GC serial number and oven type.
1. Use a Control Module to change the GC serial
number. See
"Change GC serial number"
page 1-263.
2. Use a Control Module to check/change the
oven type configuration. See
type"
on page 1-263.
3.
Released: April 2004
on
"Change oven
6850 Series Gas Chromatograph Service Procedures