Agilent Technologies 6850 Series Service Information page 234

Gc system

Advertisement

2. Remove the nut and ferrules from the coolant
inlet fitting. Use these parts to attach the
filter/connector assembly to the inlet fitting.
Released: April 2004
Close the instrument
1. Place the back panel against the GC. Recon-
2. Lower the stop plate on the lid cam. Pull the
3. Raise the stop plate and tighten the screw.
4. Connect a supply of CO
5. Restore power.
6850 Series Gas Chromatograph Service Procedures
nect the fan cable. Reinstall the seven
screws that hold the panel to the GC.
lid forward until the cam follower rests on
the curved surface of the cam in front of the
stop plate.
Close the lid.
the filter. See the next section for details.
to the free end of
2
235