9. Lift the inlet assembly and module out of the
lid. Remove the remaining insulation from the
hole under the inlet.
10. Unplug the heater/sensor leads from their
connector to the left of the inlet.
11. Lift the inlet out of the instrument.
Figure 1-15. Removing the top inlet plate
Released: April 2004
12. If you need to remove the thermal block,
Figure 1-16. Removing the thermal block
6850 Series Gas Chromatograph Service Procedures
slide off the insulation sleeve and use a
wrench to remove the bottom nut. Slide the
thermal block off the inlet weldment.
Inlet weldment
Thermal block
Bottom nut
70