Agilent Technologies 6850 Series Service Information page 161

Gc system

Advertisement

Microcell electron capture
detector
Entire detector/detector cell
WARNING
Before proceeding, turn off the oven and any
heated zones and let them cool. Turn off any
detector gases at their supply, then turn off
the main power switch and unplug the power
cord.
Released: April 2004
1. Remove the detector cover, the electronics
2. Disconnect the electrometer ribbon cable
6850 Series Gas Chromatograph Service Procedures
WARNING
The ECD micro-cell contains radioactive
To reduce the risk of exposure, wear dispos-
able gloves while handling the ECD micro-
cell. When you are finished, dispose of the
gloves and wash your hands with soap and
water.
carrier cover and the right side cover.
Caution
Make sure you are properly grounded with an
ESD strap before continuing.
from the µECD interface card.
63
Ni.
161