The µECD detector is now properly disassembled
for replacement or exchange. DO NOT remove the
inner thermal cover if you are replacing the entire
detector assembly.
Heater/Sensor assembly
After removing the µECD detector from the GC,
you can further disassemble it to replace the
heater/sensor assembly.
WARNING
The ECD microcell contains radioactive
To reduce the risk of exposure, wear dispos-
able gloves while handling the ECD micro-
cell. When you are finished, dispose of the
gloves and wash your hands with soap and
water.
Released: April 2004
1. Remove the detector as described in the
2. Loosen the locking tab screw on top of the
63
Ni.
Figure 1-11. Unlocking the thermal cover
6850 Series Gas Chromatograph Service Procedures
"Entire detector/detector cell"
page 1-161.
detector, slide the locking tab back, and
pivot it out of the away.
on
166