8. Slide the AC board assembly from the GC
chassis.
To install the AC board
Caution
If the voltage configuration is not set
correctly on the new AC board, or if incor-
rect fuses are installed, your instrument can
be damaged.
1. Compare the voltage configuration plugs on
the old and new boards (connector P12). If
necessary, transfer the old configuration
jumper onto the new board. See
and
Table
1-3.
Released: April 2004
Table 1-2. AC Board Fuses
Description
F1, F2
F3, F4
Table 1-2
6850 Series Gas Chromatograph Service Procedures
Part number Rating and type
2110-0054
6A 250 VAC IEC type F
(non time-delay), glass
body
2110-0056
15A 250 VAC IEC type F
(non time-delay), ceramic
body