Gc system

Advertisement

3. Disconnect the heater/sensor leads from the
connector on the right side of the GC.
Figure 1-8. Disconnecting the µECD cables
Released: April 2004
4. Inside the oven, remove the insulation cup
Need New
5. Use a 9/16-inch wrench to loosen the
Heater/sensor
cable
Ribbon cable
connector
6850 Series Gas Chromatograph Service Procedures
and disconnect the column from the makeup
gas adapter.
1/4-inch Swagelok nut on the makeup gas
adapter from the bottom of the detector.
Slide the makeup gas adapter out of the
bottom of the detector.