7. If the split vent trap cartridge filter was
replaced, confirm that both O-rings were
installed correctly.
O-rings
Exploded split vent trap cartridge
Released: April 2004
8. If the insert weldment gang fitting has been
6850 Series Gas Chromatograph Service Procedures
removed from the EPC module, check that
the correct length screw was used to re-
attach. If the screw is too long, the fitting
will not seal.
Split vent
trap
cartridge
Split vent
tubing
Gang
fitting
280