Rail Cam
Replacement Procedure
1.
Remove screw from lever release.
2.
Remove screw from lever rail cam.
3.
Remove spring from the lever rail cam.
4.
Reassemble in reverse order.
A
Rail Cam Kit
C
Rail Cam only Kit
Item
Description
a
Alaris GP Docking Station Kit
b
Alaris GP Fastener Spares Kit
c
Alaris SP Cam Rail Clamp Only Kit
B
Screw 70cNm
B
Screw 70cNm
A
1000SM00013 Iss. 4
Rail Cam Kit
Part Number
1000SP01307
1000SP01252
1000SP01323
55/80