Display PCB
Replacement Procedure
1.
Remove the two fixing screws from display frame.
2.
Remove Display Frame and Gasket as required.
3.
Reassemble in reverse order.
C
Screw (x2) 65cNm
A
Display
B
Display Frame
Item
Description
a
Alaris GP Display Kit
b
Alaris GP Display Accessories Kit
c
Alaris GP Fastener Spares Kit
1000SM00013 Iss. 4
Alaris® GP Volumetric Pump
Corrective Maintenance
Part Number
1000SP01255
1000SP01297
1000SP01252
63/80