CareFusion Alaris GP Technical & Service Manual page 20

Volumetric pump
Hide thumbs Also See for Alaris GP:
Table of Contents

Advertisement

Battery Calibration
Select the required option using the
1.
Connect AC Mains to the Pump and press
2.
When calibration is complete it will display
CALIBRATION FAILURE. Press PASS to confirm successful calibration or REPEAT
to perform calibration again.
BATTERY CALIBRATION
Serial No
Current
FCC
Chrg Remain
Rel Chrg
Temperature
Calibration
CAL to calibrate!
CAL can exceed 10 hrs!
QUIT
DETAILS
BATTERY CALIBRATION
Serial No
Current
FCC
Chrg Remain
Rel Chrg
Temperature
Calibration
CALIBRATION SUCCESS
QUIT
REPEAT
f
keys and the OK softkey.
CAL to begin.
CALIBRATION SUCCESS or
12345
0mA
2502mAh
2453mAh
98%
22°C
CAL
12345
0mA
2335mAh
2332mAh
100%
26°C
PASS
1000SM00013 Iss. 4
BATTERY CALIBRATION
Serial No
12345
Current
-340mA
FCC
2502mAh
Chrg Remain
2453mAh
Rel Chrg
Temperature
Calibration
CALIBRATING . . .
Discharge
QUIT
DETAILS
BATTERY CALIBRATION
Serial No
12345
Current
FCC
2502mAh
Chrg Remain
2453mAh
Rel Chrg
Temperature
Calibration
CALIBRATION FAILURE
QUIT
REPEAT
20/80
Alaris® GP Volumetric Pump
Configuration and Calibration
CALIBRATION
Zero Point Cal
Pressure Cal
Volumetric Cal
Battery Cal
SELECT WITH
QUIT
98%
22°C
89%
CAL
0mA
100%
22°C
X
OK

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents