CareFusion Alaris GP Technical & Service Manual page 42

Volumetric pump
Hide thumbs Also See for Alaris GP:
Table of Contents

Advertisement

Code
Module
MMI2
MMI3
MMI4
MMI5
MMI6
MMI7
MMI8
MMI9
MMI10
MMI11
MMI12
MMI13
MMI14
MMI15
POW1
Power Monitor
POW2
PRG1
Program
PRG2
PRG3
PRG4
PRG5
PRG6
REM1
Remote Comms
RTC1
Instrument
RTC2
SCM1
USP1
UpstreamPressure
USP2
Failure
Stuck Key Fault_Stop
Stuck Key Fault_Start
Stuck Key Fault_OnOff
Stuck Key Fault_IncInc
Stuck Key Fault_Inc
Stuck Key Fault_Dec
Stuck Key Fault_DecDec
Stuck Key Fault_Menu
Stuck Key Fault_Bolus
Stuck Key Fault_Soft1
Stuck Key Fault_Soft2
Stuck Key Fault_Soft3
Stuck Key Fault_Mute
Stuck Key Fault_Pressure
Battery Fault
Charge Fault
Flow Control Fault
Abort Fault (Prg)
Abort Fault (data)
Critical Data Corruption Fault
Image Corruption
Assertion Fault
Nurse Call Failure
RTC Init Failure
RTC Overflow Imminent
Pump Crisis
Sensor Fault
Calibration Fault
1000SM00013 Iss. 4
Alaris® GP Volumetric Pump
Action/Replace
Key has been registered as stuck for 2 minutes.
Check keypad operation in Service Mode.
Keypad or Control PCB.
Battery or Control PCB.
SD Card, Control PCB or Interface PCB.
Check Comms connections.
Comms PCB or Control PCB.
Configure clock.
Perform cold start.
Control PCB.
Switch pump off and then back on.
Contol PCB or Interface PCB.
Upstream pressure sensor, Interface PCB or
cable.
Calibrate pressure.
SD Card, Upstream pressure sensor, Interface
PCB or cable.
42/80