Battery Test and Replacement

To test the battery perform the battery calibration, as outlined in the procedure in Chapter 2 'Configuration and Calibration' , and verify
that all pass criteria are met. If pass criteria are not met then replace the battery.
Battery charge retention will eventually degrade. So where retention is critical the internal battery should be replaced every three years.

Replace the Main Battery

1.
Remove the two case screws in battery cover, remove cover and battery.
2.
Fit new battery.
3.
Replace battery cover and secure with 2 screws.
It is essential that the battery pack is calibrated after fitting as failure to do so will degrade the quoted auxiliary
A
battery power on this product .
Battery pack
Battery Cover
The battery pack used in this Alaris® Volumetric Pump is manufactured by CareFusion and includes a proprietary
A
PCB (printed circuit board) designed specifically for the Alaris® Volumetric Pump, and in conjunction with Alaris®
Volumetric Pump software, controls battery use, charge and temperature . Any use of battery packs that are not
manufactured by CareFusion in the Alaris® Volumetric Pump is at your sole risk, and CareFusion does not provide
any warranty for or endorsement on any battery packs that are not manufactured by CareFusion . CareFusion's
product warranty shall not apply in the event the Alaris® Volumetric Pump has suffered damage or premature wear,
or malfunctions or otherwise operates incorrectly, as a result of use with a battery pack that is not manufactured by
CareFusion .
Screw/Washer (x2) 40cNm
1000SM00013 Iss. 4
Alaris® GP Volumetric Pump
Preventative Maintenance