Updates; Upgrading Firmware - CareFusion Alaris GP Technical & Service Manual

Volumetric pump
Hide thumbs Also See for Alaris GP:
Table of Contents

Advertisement

Updates

Upgrading firmware

Recommended at the next service: If the Alaris® GP Guardrails® Volumetric Pump (SKU: 80263UN01-G) has
software versions V1 .9 .0 then upgrade to software version V1 .9 .2 or greater .
Mandatory: If the Alaris® GP Volumetric Pump (SKU: 80263UN01 or 80063UN01) has software versions V1 .7 .8 or
earlier then upgrade to software version V1 .7 .18 .
Mandatory: If the Alaris® GP (Guardrails®) Volumetric Pump (with Plus software) (SKU: 9002MED01 or
9002MED01-G) has software versions V2 .1 .12 or earlier then upgrade to software version V2 .1 .14 .
Complete and return the 'Software Upgrade Record' in the 'Appendix' section after performing any software
A
upgrade .
Requirements
PC
Minimum hardware system requirements:
1GHz Pentium processor
1GB of free space on the system hard drive
Available configurable RS232 9 pin serial or
USB communications ports
Video resolution of 1024 X 768 pixels and 16
bit colour depth
The Software Maintenance Utility (SMU) (1000CD00028)
RS232 cable (1000SP01183)
USB to RS232 Converter cable (1000EL00979) - optional
USB to RS232 converter 4 way hub (1000EL00980) - optional
Firmware CD
Pump Firmware Matrix
Software version
Standard Software
Part Number
Preparation
Load the SMU software onto the PC
Connect RS232 connector (using USB to RS232 converter if required) to each pump being updated
Disconnect the Battery
Remove the Battery Compartment Plug
Battery
Compartment
Plug
Guardrails® Safety Software
V1.7.18
1000SP01412
1000SM00013 Iss. 4
512MB RAM
CD ROM drive
Keyboard
Mouse
Plus Software
V1.9.2
1000SP01430
1000SP01529
24/80
Alaris® GP Volumetric Pump
Preventative Maintenance
Software requirements:
Microsoft Windows 2000 (service
pack 4), or XP (service pack 2)
V2.1.14