Power Supplies - CareFusion Alaris GP Technical & Service Manual

Volumetric pump
Hide thumbs Also See for Alaris GP:
Table of Contents

Advertisement

Power Supplies

Select the required option using the
AC Mains Test
Tests the AC mains removal detection.
Battery
Test the battery. To perform calibration see Chapter 2
'Configuration & Calibration' .
AC Mains Test
START softkey to begin.
1.
Press the
2.
Check AC Mains connected/disconnected is correctly indicated and press
O
softkey or press
softkey if not correctly detected.
3.
Press
PASS softkey to confirm pass and exit.
Battery
1.
Review the battery information.
2.
Press the
DETAILS softkey to see further battery details.
3.
Press
QUIT softkey to exit.
f
keys and the OK softkey.
1000SM00013 Iss. 4
Alaris® GP Volumetric Pump
POWER SUPPLIES
AC Mains Test
Battery
SELECT WITH
QUIT
AC MAINS TEST
P
AC Connection
START to begin...
QUIT
DETAILS
BATTERY CALIBRATION
Serial No
Current
FCC
Chrg Remain
Rel Chrg
Temperature
QUIT
DETAILS
31/80
Preventative Maintenance
OK
START
1001
0mA
2714mAh
2477mAh
91%
25°C

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents