CareFusion Alaris GP Technical & Service Manual page 33

Volumetric pump
Hide thumbs Also See for Alaris GP:
Table of Contents

Advertisement

Safety Clamp Test
START softkey to begin.
1.
Press the
2.
Check Clamp enabled/disabled is correctly indicated and press
O
press
softkey if not correctly detected.
3.
Press
PASS softkey to confirm pass and exit.
Sear Test
START softkey to begin.
1.
Press the
2.
Check Sear in/out is correctly indicated and press
not correctly detected.
3.
Press
PASS softkey to confirm pass and exit.
Air In Line Test
1.
Press the
START softkey to begin.
2.
Insert a fluid filled Infusion Set and an air filled Infusion Set.
3.
Confirm pump detects fluid and air correctly and press
softkey if not correctly detected.
PASS softkey to confirm pass and exit.
4.
Press
P
softkey or press
P
softkey or press
1000SM00013 Iss. 4
SAFETY CLAMP TEST
P
softkey or
Working
Clamp enabled
Test
Opto Tx En
Opto Tx Dis
QUIT
O
softkey if
Working
Sear in place
Test
QUIT
AIR IN LINE TEST
O
Upstream:
Downstream:
QUIT
33/80