Pressure Calibration
Equipment required:
Calibrated Pressure Gauge, minimum specification of
Accuracy = 0.10% of full scale
Full scale = 1500mmHg
Pressure Calibration Set (1000SP01422) Use to calibrate 10 pumps and then
change
Set up equipment as per figure 2-1 and allow 30 seconds before proceeding.
Select the required option using the
1.
Turn 3-way tap to close from atmosphere.
2.
Press the
START softkey.
3.
Apply a pressure of 200mmHg and pump will countdown for 15 seconds.
4.
Press the
ACCEPT softkey.
5.
Apply a pressure of 800mmHg and pump will countdown for 15 seconds.
6.
Press the
ACCEPT softkey.
7.
Turn 3-way tap to vent to atmosphere.
8.
Turn 3-way tap to close from atmosphere.
9.
Press the
NEXT softkey to proceed to Verification Procedure.
Verification Procedure
Software version v1.7 x and below
10.
Apply a pressure of 200mmHg and wait for 5 seconds.
NEXT softkey.
11.
Press the
12.
Apply a pressure of 400mmHg and wait for 5 seconds.
13.
Press the
NEXT softkey.
14.
Apply a pressure of 600mmHg and wait for 5 seconds.
NEXT softkey.
15.
Press the
16.
Apply a pressure of 800mmHg and wait for 5 seconds.
17.
Press the
NEXT softkey.
18.
Turn 3-way tap to vent to atmosphere.
PASS softkey.
19.
Press the
Pressure source
(50ml/100ml syringe or
similar device)
Pressure
Gauge
Figure 2 - 1 Pressure Calibration Equipment Set Up
keys and the OK softkey.
f
Vent to
atmosphere
1000SM00013 Iss. 4
CALIBRATION
Zero Point Cal
Pressure Cal
Volumetric Cal
Battery Cal
SELECT WITH
QUIT
Software version v1.9.x and above
10.
Apply a pressure of 500mmHg and wait for 5 seconds.
NEXT softkey.
11.
Press the
12.
Turn 3-way tap to vent to atmosphere.
13.
Press the
PASS softkey.
18/80
Alaris® GP Volumetric Pump
Configuration and Calibration
OK
Alar s®
GP