CareFusion Alaris GP Technical & Service Manual page 16

Volumetric pump
Hide thumbs Also See for Alaris GP:
Table of Contents

Advertisement

Upload data set to an Alaris® GP (Guardrails®) Volumetric Pump (with Plus software)
Equipment required:
Alaris® Editor V4.2 Software Kit (1000SP01462) or Alaris® Transfer Tool Software Kit (1000SP01463)
RS232 cable (1000SP01183)
USB to RS232 Converter cable (1000EL00979)
USB to RS232 converter 4 way hub (1000EL00980) - optional
PC - for requirements see Upgrading Firmware PC requirements
Using the Alaris® Transfer Tool allows an approved data set to be uploaded to a pump (with Plus software).
Warning -
A
At no time should the Alaris® Transfer Tool be used to upload to a pump while the pump is connected to a patient .
In Service Mode select Data Set Transfer using the
Data Set Upload button (Ctrl+U). The Data Set Upload Wizard is displayed. Click Next to continue.
1.
Select the
2.
Select the serial port(s) required to transfer the data set to the pump(s). Click
3.
Click the
Select Dataset button and Open the required data set. The selected data set and path are displayed. Click Next to
continue.
4.
Enter the data set ID and click
become available. Click Next to continue.
5.
The data set selection is now complete. Click
Supported Pump Types.
Connect the pump to PC.
6.
7.
Press the
START softkey to begin transfer.
8.
When the transfer is complete, the pump will validate the data then display
Data Set Deployed to the Pump.
Note: Please ensure the data set ID shown on the pump is identical to the one transferred.
9.
Press
ACCEPT softkey to confirm correct transfer and exit. The transfer status will be shown as Deployed.
10.
Switch off the pump by holding the power button for 3 seconds and disconnect from the PC.
11.
If multiple uploads are required, connect the next pump and repeat from step 5.
12.
When all uploads are complete, click
Caution: Loading the Transfer Tool software is considered a non-clinical service activity . Interconnecting the pump
A
with a PC may cause the safety or electromagnetic environment to change while the connection exists . The threat of
higher leakage currents or EMI disturbances may be present . Disconnect the IrDA or RS232 cable connection at both
ends following software upload activities .
Note: For more
information relating to
the Alaris® Editor Software
refer to the relevant
Directions For Use supplied
with the software.
f
keys and the OK softkey.
Validate. If the data set ID entered matches the ID of the selected data set the Next button will
Finish. The file information is updated to show the File used, Data Set and
Menu > Exit to close the Transfer Tool.
1000SM00013 Iss. 4
Alaris® GP Volumetric Pump
Configuration and Calibration
Next to continue.
Transfer Complete. The transfer status will be shown as
16/80