General Fault Diagnosis; Exception Error Handling - CareFusion Alaris GP Technical & Service Manual

Volumetric pump
Hide thumbs Also See for Alaris GP:
Table of Contents

Advertisement

General Fault Diagnosis

Failure
Display missing vertical lines
No response from keypad or LED
Safety alarm is activated
Failure of RS232 communications
Dropped or damaged
Exposed to fluids
No battery power
No AC mains power
Delivery rates out of
tolerance

Exception Error Handling

Exception errors include Assertion Errors and are used to trap logical errors in the software execution.
The pump will display the error type, the title of the software module in which the error occurred and the line number. The user should
make a note of these for use in diagnosis. This information is stored in the event log.
After an error, the pump will not store information when powered down. When the pump is switched on again, the user should always
confirm clear setup.
Action/Replace
If Control PCB is issue 9 and below, then replace with latest issue Control PCB.
Replace Comms PCB.
1000SM00013 Iss. 4
Parts to Check/Test
43/80
Alaris® GP Volumetric Pump
Troubleshooting