Alaris® GP Volumetric Pump
General Information

General Precautions

w
Attention consult accompanying documents: Prior to using this pump, carefully read the Operating Precautions
described in the Directions for Use (DFU).
V
This pump contains static-sensitive components. Observe strict precautions for the protection of static sensitive
components when attempting to repair and service the pump.
B
An explosion hazard exists if the pump is used in the presence of flammable anaesthetics. Exercise care to locate
the pump away from any such hazardous sources.
m
Dangerous Voltage. An electrical shock hazard exists if the casing of the pump is opened or removed. Refer all
servicing to qualified service personnel.
This pump is protected against the effects of high energy radio frequency emissions and is designed to be fail
M
safe if extremely high levels of interference are encountered. Should false alarm conditions be encountered,
either remove the source of the interference or regulate the infusion by another appropriate means.
If the pump is dropped, subjected to excessive moisture, humidity or high temperature, or otherwise suspected
to have been damaged, remove it from service for inspection by qualified service personnel.
When connected to an external power source, a three-wire (Live, Neutral, Earth) supply must be used. If the
d
integrity of the external protective conductor in the installation or its arrangement is in doubt, the pump should
be operated from the battery.
1000SM00013 Iss. 4