Control PCB and Interface PCB
Replacement Procedure
1.
Disconnect display and keypad cables from Control PCB.
2.
Remove the three retaining screws and washers.
3.
Remove the three pillar supports.
4.
When fitting Control PCB ensure all flexi and cables are routed clear of PCB.
5.
Reassemble in reverse order.
Recommended when serviced: Replace the SD card (1000SP01299) for Alaris® GP Volumetric Pumps with serial
numbers between 802608618 and 802611376 that had not exhibited a DSC2 fault prior to June 2008 .
Item
Description
a
Alaris GP Interface PCB Kit
b
Alaris GP Control PCB Kit
c
Alaris GP Fastener Spares Kit
Note: Three Pillar Supports are supplied with both PCB kits.
A
Interface PCB
C
Screw (x3)
65cNm
1000SM00013 Iss. 4
Alaris® GP Volumetric Pump
Corrective Maintenance
B
Control PCB
Part Number
1000SP01300
1000SP01298
1000SP01252
62/80
Pillar supports (x3)