Handle - CareFusion Alaris GP Technical & Service Manual

Volumetric pump
Hide thumbs Also See for Alaris GP:
Table of Contents

Advertisement

Handle

Replacement Procedure
1.
Remove two screw from handle block.
2.
Remove handle block, handle spring and handle.
3.
Reassemble in reverse order.
Refitting notes:
Make sure that the handle spring is in front of the handle and not behind it.
Item
Description
a
Asena LVP Overmould Handle
b
Handle Spring Asena LVP
c
Asena LVP Handle Retaining Block
d
Alaris GP Fastener Spares Kit
1000SM00013 Iss. 4
Alaris® GP Volumetric Pump
D
Screw (x2) 40cNm
C
Handle block
B
Handle spring
A
Handle
Part Number
1000ME01845
1000ME00630
1000ME00632
1000SP01252