Advertisement

Data Set Transfer

Upload data set to an Alaris® GP Volumetric Pump
Equipment required:
Alaris® GP Editor Software Kit (1000SP01310) - includes the Alaris® GP Transfer Tool
RS232 cable (1000SP01183)
USB to RS232 Converter cable (1000EL00979) - optional
USB to RS232 converter 4 way hub (1000EL00980) - optional
PC - for requirements see Upgrading Firmware PC requirements
Using the Alaris® GP Transfer Tool allows a released data set to be uploaded to an Alaris® GP Volumetric Pump.
Warning -
A
At no time should the Alaris® GP Transfer Tool be used to upload to an Alaris® GP Volumetric Pump while the pump is
connected to a patient .
In Service Mode select Data Set Transfer using the
1.
Using the Alaris® GP Transfer Tool select data set to be uploaded.
2.
Press the
RS232 or IrDA softkey to select the Comms mode being used.
3.
Connect the pump to PC.
START softkey to begin transfer.
4.
Press the
5.
Please ensure the data set ID shown on the pump is identical to the one transferred.
6.
Press
PASS softkey to confirm correct transfer and exit.
7.
To transfer the data set to another pump repeat steps 2 to 6.
Upload data set to an Alaris® GP Guardrails® Volumetric Pump
Equipment required:
Guardrails® Editor V3.1 Software Kit (1000SP01389) or Guardrails® Editor V3.1 Transfer Tool Software Kit (1000SP01390)
RS232 cable (1000SP01183)
USB to RS232 Converter cable (1000EL00979)
USB to RS232 converter 4 way hub (1000EL00980) - optional
PC - for requirements see Upgrading Firmware PC requirements
Using the Guardrails® Editor V3.1 Transfer Tool allows an approved data set to be uploaded to an Alaris® GP Guardrails® Volumetric Pump.
Warning -
A
At no time should the Guradrails® Editor V3 .1 Transfer Tool be used to upload to an Alaris® GP Guardrails® Volumetric
Pump while the pump is connected to a patient .
In Service Mode select Data Set Transfer using the
1.
Using the Guardrails® Editor V3.1 Transfer Tool Transfer Tool select data set to be uploaded.
2.
Connect the pump to PC.
START softkey to begin transfer.
3.
Press the
4.
Please ensure the data set ID shown on the pump is identical to the one transferred.
5.
Press
ACCEPT softkey to confirm correct transfer and exit.
6.
To transfer the data set to another pump repeat steps 2 to 5.
Caution: Loading the Data Set Transfer Tool software is considered a non-clinical service activity . Interconnecting the
A
pump with a PC may cause the safety or electromagnetic environment to change while the connection exists . The
threat of higher leakage currents or EMI disturbances may be present . Disconnect the IrDA or RS232 cable connection
at both ends following software upload activities .
Note: For more
information relating to the
Alaris® GP Editor Software
or the Guardrails® Editor
V3.1 Software refer to the
relevant Directions For Use
supplied with the software.
f
keys and the OK softkey.
keys and the OK softkey.
f
1000SM00013 Iss. 4
Alaris® GP Volumetric Pump
Configuration and Calibration
15/80