RS232 Connector and Comms PCB

Replacement Procedure
1.
Remove two retaining screws and washers from assembly.
2.
Remove RS232 connector cover and two RS232 socket screws.
3.
Remove RS232 Connector & Comms PCB assembly.
4.
Reassemble in reverse order.
A
Comms PCB
Item
Description
a
Alaris GP Comms PCB Kit
b
Cover RS232
c
Alaris GP Fastener Spares Kit
d
Asena GW, Assy, Cover Dust Drop Sensor
e
Fuse Cover
C
Screw/washer 40cNm
C
Screw 40cNm
B
Connector cover
Connector cover
1000SM00013 Iss. 4
C
Screw (x2) 40cNm
RS232
D
Flow Sensor
A
RS232 Socket
Screw (x2) 40cNm
Part Number
1000SP01256
1000ME01745
1000SP01252
1000ME00291
1000ME00655
56/80
Alaris® GP Volumetric Pump
Corrective Maintenance
C
Screw (x2) 40cNm