Calibration - CareFusion Alaris GP Technical & Service Manual

Volumetric pump
Hide thumbs Also See for Alaris GP:
Table of Contents

Advertisement

Calibration

Zero Point Calibration
Select the required option using the
1.
Wait for the pressure sensors to park.
2.
Ensure that an infusion set is not installed.
START softkey.
3.
Press the
4.
Pump will countdown for 15 seconds.
5.
Press the
ACCEPT softkey.
6.
Press the
PASS softkey.
f
keys and the OK softkey.
1000SM00013 Iss. 4
Alaris® GP Volumetric Pump
Configuration and Calibration
CALIBRATION
Zero Point Cal
Pressure Cal
Volumetric Cal
Battery Cal
SELECT WITH
QUIT
17/80
OK