Software Upload
1.
Through the Rear Case, there is a set of three dip switches on the bottom of the Control PCB. Switch 1 should be ON and switches 2
and 3 should be OFF.
2.
Using a small flat blade screwdriver or round ended tweezers configure the dip switch settings
to:
a)
Switch 1 = OFF
b)
Switch 2 = ON
c)
Switch 3 = ON
3.
Start the SMU facility to upload the software.
4.
Select the Firmware zip file.
5.
Select the Comm settings - Comm Port and Baud rate of 460800 (USB to RS232 converter) or
115200 (RS232).
6.
Connect the Battery to the Battery Cable.
7.
Press
Start.
8.
Once the green bar has reached the far right hand side and the time has reached 0:00 and
the flashing green light is a steady green light, the RS232 connector can be removed from the
pump.
9.
Disconnect the Battery and turn OFF all dip switches.
10.
Wait 5 seconds then reconnect the Battery.
11.
Configure the Dip Switches to:
a)
Switch 1 = ON
b)
Switch 2 = ON then OFF
c)
Switch 3 = OFF
12.
Refit the Battery Compartment Plug, this prevents fluid ingress.
13.
Power up the pump in Service Mode, enter access code
time.
If the Control, Interface or RS232 PCB is replaced, the pump must be re-programmed .
A
212, then select Configuration > Date/Time and set the current date and
1000SM00013 Iss. 4
Alaris® GP Volumetric Pump
Preventative Maintenance
Switch 1 turns the safety
battery circuit on but is not
required for programming.
Switch 2 forces the pump
to turn on.
Switch 3 turns the
pump into 'Boot Mode' ,
this is only used for
programming.