CareFusion Alaris GP Technical & Service Manual page 11

Volumetric pump
Hide thumbs Also See for Alaris GP:
Table of Contents

Advertisement

Clear CQI Log File
OK to confirm clearing the CQI Log File.
Press
Configuration
Select the required option using the
Date/Time
Sets the current date and time used for event logging.
Software Versions
Displays the pump software versions.
Serial Number
Configure the displayed serial number.
Pump Reference
Pump specific text to be displayed in user mode at start
up. (20 characters max.)
Language
Configure the Language used for display messages.
Backlight & Contrast
Adjust the Backlight and Contrast values
Current Data Set File
Displays the current data set file details.
Date/Time
1.
Set the correct date and time using the
NEXT to continue to next item to change.
2.
Press
3.
Press
OK to confirm.
f
keys and the OK softkey.
f
keys.
1000SM00013 Iss. 4
Alaris® GP Volumetric Pump
Configuration and Calibration
CLEAR CQI LOG FILE
******
WARNING!
This will remove all
instrument history and will
restore the factory default
data set by deleting the
clinically approved
installed data set.
Clear current log?
QUIT
SERVICE CONFIGURATION
Date/Time
Software Versions
Serial Number
Pump Reference
Language
Backlight & Contrast
Current Data Set File
SELECT WITH
QUIT
DATE / TIME
AUG–04–2006
09:48
ADJUST WITH
QUIT
NEXT
11/80
******