Alarms Functionality - CareFusion Alaris GP Technical & Service Manual

Volumetric pump
Hide thumbs Also See for Alaris GP:
Table of Contents

Advertisement

Alarms Functionality

1.
Press the
START softkey to begin.
2.
Pump starts an infusion.
3.
Check all alarms sound and display correctly.
4.
Alarms to test are:
AC Mains Disconnect
Door Open
Upstream occlusion
Air In Line
5.
When all the alarms have been tested press the
to be checked.
ALARMS FUNCTIONALITY
****
INFORMATION!
Ensure that a fluid filled
set is in use.
Subsequent alarm
screens should be
cleared using the
CANCEL softkey.
QUIT
PASS softkey if all the alarms worked correctly or REPEAT softkey if alarms still need
*****
START
ALARMS FUNCTIONALITY
AC Mains
Door Open
Upstream Occ
Air In Line
All alarms detected
QUIT
REPEAT
1000SM00013 Iss. 4
ALARMS FUNCTIONALITY
AC Mains
Door Open
Upstream Occ
Air In Line
Waiting for alarm . . .
QUIT
PASS
28/80
Alaris® GP Volumetric Pump
Preventative Maintenance
.
.
.
.
STOP

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents