Performance Verification Procedure

Model / Serial Number:
Hospital Name / Reference:
Inspection
Physical inspection and clean
Check all functions in PVP Work Flow
Enter access code 212 and go to PVP Work Flow
CH3
Self Test
Setup
Set rate to zero (or lowest value possible), clear Volume Infused and VTBI
Class I Type CF
Test in accordance with the standard
EN 60601-1 and test equipment
Electrical Safety
operation manual.
TestS
Verification
Performed
By
____________________________
Sign
CHX indicates the chapter number in the Technical Service Manual (TSM) - 1000SM00013.
E.G. CH2 = Refer to TSM Chapter 2.
CH3
Software Versions
Date/Time
Serial Number
Door Frame
Safety Clamp
Sear
Audio Test
Chequerboard
LED's
Keypad
Nurse Call
RS232 Loop Back
Pumping Efficiency Test
-
Time taken = 2 minutes or less
Downstream Occlusion Pressure Test
-
Occlusion alarm = 500 ± 100 mmHg
Alarms Functionality
Volumetric (Accuracy Test) Calibration
-
Delivery = 20 ml ± 0.6 ml (3%)
1000SM00013 Iss. 4
Service Order / Inventory Number:
Software Version:
Test results are stored:
Electronically
¨
Print-out
¨
Other
¨
____________________
____________________________
Print
39/80
Alaris® GP Volumetric Pump
Preventative Maintenance
______ minutes ______seconds
______________mmHg
______________ml
PASS / FAIL