Recommended Cleaning and Storage
Cleaning the Pump
Before the transfer of the Pump to a new patient and periodically during the use, clean the Pump by wiping over with a lint-free cloth
lightly dampened with warm water and a standard disinfectant / detergent solution.
Do not use the following disinfectant types:
•
NaDcc (such as PRESEPT)
•
Hypochlorites (such as CHLORASOL)
•
Aldehydes (such as CIDEX)
•
Cationic Surfactants (such as Benzalkonium Chloride)
•
Iodine (such as Betadine)
•
Concentrated Isopropyl alcohol based cleaners will degrade plastic parts.
Recommended cleaners are:
Brand
Concentration
Hibiscrub
20% (v/v)
Virkon
1% (w/v)
The following products were tested and are acceptable for use on the Alaris® GP Volumetric Pump range if used in accordance with the
specified manufacturer's guidelines.
•
Warm soapy water
•
Mild detergent in water (e.g. Young's Hospec)
•
70% Isopropyl Alcohol in water
•
Chlor-Clean
•
Clinell Sporicidal wipes
•
Hibiscrub
•
TriGene Advance
•
Tristel Fuse sachets
•
Tristel Trio wipes system
•
Tuffie 5 wipe
•
Virkon Disinfectant
•
Virusolve+ (Ready To Use)
•
Virusolve+ (Wipes)
Before cleaning always switch OFF and disconnect from the AC power supply . Never allow fluid to enter the casing
A
and avoid excess fluid build up on the Pump .
Do not use aggressive cleaning agents as these may damage the exterior surface of the Pump .
Do not steam autoclave, ethylene oxide sterilise or immerse this Pump in any fluid .
1000SM00013 Iss. 4
Alaris® GP Volumetric Pump
Preventative Maintenance
22/80