Mains inlet, IrDA PCB, PE stud and magnet

Replacement Procedure
1.
Remove nut and washer to remove PE stud.
2.
Remove the two screws on Mains inlet retaining plate.
3.
Remove mains inlet retaining plate.
4.
Remove magnet by lifting one end.
5.
Unclip IrDA PCB and remove.
6.
Unclip Mains Inlet and remove.
7.
Reassemble in reverse order.
E
PE Stud
Item
Description
a
Alaris GP Mains Inlet Kit
b
Alaris GP IrDA PCB Flexi Kit
c
Alaris GP Fastener Spares Kit
d
Magnet IR Detect
e
Stud PE Connector M6 Thread X 15
f
Asena LVP Mains Inlet Bracket
*
Bussmann Fuse Gmd-1.25A
D
1000SM00013 Iss. 4
AF
Mains Inlet
A
AC Mains
and Bracket
Inlet Gasket
C
Washer
Magnet
C
150cNm
Part Number
1000SP01251
1000SP01308
1000SP01252
1000ME01303
0000ME00141
1000ME00636
0000ME00770
53/80