Volumetric Calibration
Select the required option using the
1.
Load the primed Infusion Set (60793) into the Pump and set-up as shown in
Figure 2-2 below and adjust the fluid level so that the meniscus is level with the
zero mark.
2.
Press
START to begin. Test will run and fluid will be delivered into the burette.
3.
When
Volume delivered! is displayed, check accumulated air in line value
is less than 100µl then enter the volume delivered into the burette using the
f
keys and the OK softkey. If accumulated air in line value is greater
than 100µl then repeat test.
4.
If no calibration is required (Volume delivered within limits of 19.4ml to 20.5ml)
then press PASS to confirm and exit.
5.
If the Calibration value is changed automatically then press
steps 2 to 4.
6.
If the pump still fails replace the Platen and Fingers then repeat the calibration
procedure.
Figure 2 - 2 Volumetric Calibration Equipment Set Up
f
keys and the OK softkey.
VERIFY and repeat
Al r s
GP
1000SM00013 Iss. 4
CALIBRATION
Zero Point Cal
Pressure Cal
Volumetric Cal
Battery Cal
SELECT WITH
QUIT
VOLUMETRIC CALIBRATION
Rate
VTBI
Cal Value
Acc'd Air
450mm
± 20mm
QUIT
19/80
Alaris® GP Volumetric Pump
Configuration and Calibration
OK
125ml/h
20.0ml
920
0µl
START to begin...