Power Supply; Iec Mains Inlet; Membrane Keypad - CareFusion Alaris GP Technical & Service Manual

Volumetric pump
Hide thumbs Also See for Alaris GP:
Table of Contents

Advertisement

Alaris® GP Volumetric Pump
Circuit Descriptions

Power Supply

A universal input switched mode power supply used to regulate the mains input voltage.

IEC Mains Inlet

A medical grade filtered mains inlet with fuses in the live and neutral lines. The fuses can be accessed by removing the external splash
cover and opening the fuse draw.

Membrane Keypad

The membrane keypad consists of fourteen keys and LED's to indicate battery, mains, start and stop. The on / off key connects to the
main processor and the safety processor. The safety processor manages the power up sequence and the main processor power down.
1000SM00013 Iss. 4