11.2.2 The Accuracy at the Computation

(1) Scaling

The accuracy rating at the scaling is as following formula.
Scaling accuracy (digits) = Range accuracy rating (digits) × Scaling factor +2 digits
But, Scaling factor=
Example 1)When VOLT range is -1.000~1.000V and Scaling range is 0.00~100.00.
Range accuracy rating =0.1%×1.000V+3 digits
(Input +1.000V) =4 digits
Scaling factor
Therefore, Scaling accuracy
Analog recording accuracy =± 4+{1000-(-1000)}×
Example 2)When Type K is 0.0~200.0℃ and Scaling is 0.00~200.00.
Range accuracy rating
(Input 200℃)
Scaling factor=
Therefore, Scaling accuracy

(2) Square root computation

The square root computation is as following formula.
Display value=10
Display value accuracy rating:ε, Process variable accuracy: e, and
Process variable: x (convert into %).
=10
Display value
Therefore, Display value accuracyεε=
Scaling span (digits)
Measurement range span (digits)
10000-0
1000-(-1000)
=±(4×5+2)
=±22 digits
=±10 digits
=±0.01V
=0.15%×200.0+0.3℃
=0.6℃
=6 digits
20000-0
=10
2000-0
=±(6×10+2)
=±62 digits
=±0.62℃
Process variable
x ±
x±e
10
5e
―124―
=5
0.3
100
10e
2 x
x
HXPRM10mnCT002E