Digital Filter; Settings Related To Recording; Recording Start/Stop Trigger Setting - Toho Electronics TRM-10C Instruction Manual

Hybrid pen type recorder
Hide thumbs Also See for TRM-10C:
Table of Contents

Advertisement

7.2.5 Digital Filter

Apply a digital filter to a process valuable to record the first order lag. From 99.5% response time
(T
), the filter constant (K) is obtained by the following formula.
0.64
K=
When 99.5% response time is 50 second; K=
Example)
Set up the digital filter constant for Channel 2 to "K = 0.0128".
Display

7.2.6 Settings Related to Recording

Sets a recording start/stop trigger, select tag/channel print, or set alarm print and logging print.

(1) Recording start/stop trigger setting

Set whether the trigger of recording start/stop should be by the "RUN" key or DI.
Example)
When setting DI as the recording start/stop trigger
Display
[Note]
When indicates "
in page 89
.
Operation Keys
ENT
Enter the engineering mode (See the key operation
in page 78).
ENT
PRINT
Use the △ key to select
PRINT
ENT
"ENT" key.
Use the △ key to alter a numerical value, or the
PRINT
FEED
key to shift a digit. Press the "ENT" key.
ENT
Setting is completed.
PRINT
ENT
setting.
Operation Keys
ENT
Enter the engineering mode (See the key operation
in page 78). Use the △ key to indicate "
ENT
PRINT
and press the "ENT" key.
Use the △ key to select "
ENT
PRINT
"ENT" key.
Use the △ key to indicate "
PRINT
ENT
"ENT" key.
Setting is completed. Returned to the Recording
Start/Stop Setting screen.
", RUN/STOP key does not operate. Set "
―84―
0.64
=0.0128
50
Description
and press the
Returned to the channel
Description
", and press the
", and press the
to DI. See DI functions
HXPRM10mnCT002E
",