4.7

Block Sampling Settings

Procedure
KEY
LOCK
SEC RANGE
Explanation
IM 704610-01E
Display 2
Display 1
ns ms
s
%
V
%/
S
Arming must be set to external arming before carrying out the procedures in this
section. For the procedure to set arming, see section 4.6.
1.
Press SHIFT+INHIBIT (BLOCK). Block appears in Display 2, and the setting
value appears in Display 1.
2.
Use rotary knob & < > to set the number of blocks. If the number of blocks is
set to 2 or more, the BLOCK indicator illuminates.
With block sampling, all data collected after performing one block of measurement a
specified number of repetitions is processed and displayed together. When using
external arming, you can set the number of block in the following range.
Setting range: 1 to 99 (in steps of 1 block)
Note
During normal measurement, the number of blocks is set to 1. For block sampling
measurements other than normal measurement, select a number of blocks of 2 or more.
The setting for the gate time and number of blocks is restricted such that maximum
number of samples that can be acquired by the instrument is not exceeded.
Conditional expressions
When D-to-C high speed calculation (see section 4.9) is OFF:
Gate time × no. of blocks ≤ 5 seconds
When D-to-C high speed calculation is ON: Gate time × no. of blocks ≤ 1 seconds
Example
Gate time
1 ms
10 ms
50 ms
100 ms
500 ms
1000 ms
Once the gate time and number of blocks reaches the limit according to the condition
equation above, if the gate time is increased, the setting for the number of blocks is
automatically adjusted. For example, when D-to-C high speed calculation is OFF, and
when the gate time starts out at 50 ms and the number of blocks is 99, if the gate time
is increased to 500 ms, the number of blocks automatically changes to 10. However, if
the gate time is then returned to 50 ms, the number of blocks stays at 10.
If the gate time is decreased and the measurable number of blocks increases, the
setting for the number of blocks is not automatically increased. To manually increase
the number of blocks, perform the procedure described in this section.
<<For a functional description, see section 2.4>>
L.MARK
R.MARK
DISP OFF
MEASURE
E2T
JUDGE
PW
GO
BD x1
D to C
NO-GO
GATE TIME
DELAY
BLOCK
MEDIA
FUNCTION
ARMING
INHIBIT
SHIFT
SHIFT
UTILITY
REMOTE
STORE
10%
LOCAL
RECALL
ENTER
20%
JDG LEVEL
LEVEL
COUPLE
EQ BOOST
S
POLARITY
TRIG
IMPEDANCE EQUALIZER
AUTO
MAN
50 Ω
1M Ω
CONV
SEC RANGE
SCALE
Measurable no. of blocks
When D-to-C high speed
calculation is OFF
99
99
99
50
10
5
The phrase, "rotary knob & <>"
is used in the following
explanation to instruct the user
to enter numerical setting values
using the knob and arrow keys.
For details on this procedure,
see section 3.7.
When D-to-C high speed
calculation is ON
99
99
20
10
2
1
4-11
4