13.5 Input/Output on the Rear Panel

Item
External arming signal input (EXT ARM IN)
Connector type
Input impedance
Input coupling
Input level
Allowable input voltage
range
Minimum pulse width
Inhibit signal input (INHIBIT IN)
Connector type
Input impedance
Input coupling
Input level
Allowable input voltage
range
Minimum pulse width
Jitter ratio DC output (JITTER DC OUT)
Connector type
Output impedance
Output coupling
Output mode
Output filter
Jitter ratio output range
Determination output
Correction coefficient α
Correction coefficient β
Output level
Output level accuracy
RF signal voltage level DC output (LEVEL DC OUT)
Connector type
Output impedance
Output coupling
Output mode
Output filter
Voltage level DC output
range
Determination output
Output level
Output level accuracy
RF signal monitor output (MONITOR OUT)
Connector type
Output impedance
Output coupling
‡‡†
Output level
Equalized RF signal monitor output (EQUALIZED OUT)
Connector type
Output impedance
Output coupling
‡‡†
Output level
Data signal output (SLICED RF OUT)
Connector type
Output impedance
Output coupling
‡‡†
Output level
13-4
Specifications
BNC
10 kΩ (typical value
DC
TTL level
–5 to 10 V (DC+ACpeak)
30 ns
BNC
10 kΩ (typical
)
DC
TTL level
–5 to 10 V (DC+ACpeak)
30 ns
BNC
600 Ω (typical value
DC
Select jitter ratio DC output or determination output.
Average coefficient setting range: 1 to 10, setting resolution: 1
(output filter is disabled when D-to-C high speed calculation
Setting range: 0.00 to 100.00%, setting resolution: 0.01%
Range of determination level: 0.00% to 25.00%, resolution: 0.01%
Range: 0.0001 to 9.9999, resolution: 0.0001
Range: –9.999 to 9.999%, resolution: 0.001%
0 to 5 VDC, Default setting: 0.2 V/%
‡, ‡‡
±10 mV
BNC
600 Ω (typical value
DC
Select voltage level DC output or determination output
Setting range of average coefficient: 1 to 10, setting resolution: 1
Setting range: 0.000 to 5.000 V, setting resolution: 0.001 V
Setting range of determination level: 0.000 to 5.000 V, setting resolution: 0.001 V
0 to 5 VDC, Default setting: 1 V/Vp-p
‡, ‡‡
±10 mV
BNC
50 Ω (typical value
)
DC
Approximately 1/2 the RF signal (within ±5 V)
BNC
50 Ω (typical value
)
DC
Approximately 1/2 the RF signal (within ±5 V) when the equalizer is OFF and AGC is OFF.
Approximately 0.4 to 0.7 Vp-p (within ±1 V) when the equalizer is OFF and AGC is ON.
BNC
50 Ω (typical value
)
DC
Approximately ±0.4 V
)
)
)
††
is ON)