YOKOGAWA TA220 User Manual page 111

Digital jitter meter
Table of Contents

Advertisement

:INPut?
Function
Queries all settings related to the signal input
section.
Syntax
:INPut?
Example
:INPUT?
-> :INPUT:DATA:COUPLING DC
;IMPEDANCE I1M;TRIGGER
:LEVEL 100E+00;TRIGGER:MODE MANUAL
:INPut:DATA?
Function
Queries all settings related to the signal input
section (excluding the equalizer).
Syntax
:INPut:DATA?
Example
:INPUT:DATA?
-> :INPUT:DATA:COUPLING DC
;IMPEDANCE I1M;TRIGGER
:LEVEL 100E+00;TRIGGER:MODE MANUAL
:INPut:DATA:COUPling
Function
Sets the input coupling or queries the current
setting.
Syntax
:INPut:DATA:COUPling {AC|DC}
:INPut:DATA:COUPling?
Example
:INPUT:DATAL:COUPLING DC
:INPUT:DATA:COUPLING?
-> :INPUT:DATA:COUPLING DC
:INPut:DATA:IMPedance
Function
Sets the input impedance or queries the current
setting.
Syntax
:INPut:DATA:IMPedance {I50|I1M}
:INPut:DATA:IMPedance?
Example
:INPUT:DATA:IMPEDANCE I1M
:INPUT:DATA:IMPEDANCE?
-> :INPUT:DATA:IMPEDANCE I1M
:INPut:DATA:POLarity
Function
Sets the polarity of the data signal or queries
the current setting.
Syntax
:INPut:DATA:POLarity {POSitive|
NEGative|BOTH}
:INPut:DATA:POLarity?
Example
:INPUT:DATA:POLARITY POSITIVE
:INPUT:DATA:POLARITY?
-> :INPUT:DATA:POLARITY POSITIVE
Description BOTH cannot be set when the function is
PulseWidth.
:INPut:DATA:TRIGger?
Function
Queries all settings related to the trigger.
Syntax
:INPut:DATA:TRIGger?
Example
:INPUT:DATA:TRIGGER?
-> :INPUT:DATA:TRIGGER
:LEVEL 100E+00;MODE MANUAL
IM 704610-01E
:INPut:DATA:TRIGger:LEVel
Function
Sets the trigger level (slice level) or queries the
current setting.
Syntax
:INPut:DATA:TRIGger
:LEVel {<voltage>|<NRf>}
:INPut:DATA:TRIGger:LEVel?
When the function is DTOC
• <NRf>= –1000 to 1000 (in steps of 1)
When the function is PulseWidth and the
equalizer and AGC are OFF
• <Voltage> = –2.000 V to 2.000 V (in steps of
1 mV)
When the function is PulseWidth and the
equalizer is OFF or AGC is ON
• <NRf>= –1000 to 1000 (in steps of 1)
Example
:INPUT:DATA:TRIGGER:LEVEL 100
:INPUT:DATA:TRIGGER:LEVEL?
-> :INPUT:DATA:TRIGGER
:LEVEL 100E+00
Description Cannot be set or queried when the trigger mode
is AUTO.
:INPut:DATA:TRIGger:MODE
Function
Sets the trigger mode or queries the current
setting.
Syntax
:INPut:DATA:TRIGger
:MODE {AUTO|MANual|BOTH}
:INPut:DATA:TRIGger:MODE?
Example
:INPUT:DATA:TRIGGER:MODE MANUAL
:INPUT:DATA:TRIGGER:MODE?
-> :INPUT:DATA:TRIGGER:MODE MANUAL
Description • Only MANUAL can be set when the function
is PulseWidth and EQ and AGC are OFF.
• The trigger mode is forcibly set to AUTO when
the limit equalizer is selected (see the option
function user's manual, IM704610-51E). The
trigger mode remains the same even if the
equalizer is returned to OFF or CONV.
:INPut:EQ?
Function
Queries all settings related to the equalizer.
Syntax
:INPut:EQ?
Example
:INPUT:EQ?
-> :INPUT:EQ:STATUS OFF;AGC OFF
;DCCLAMP OFF;PLLHOLD OFF
:INPut:EQ:STATus
Function
Sets the equalizer status or queries the current
setting.
Syntax
:INPut:EQ:STATus {OFF|CONV}
:INPut:EQ:STATus?
Example
:INPUT:EQ:STATUS OFF
:INPUT:EQ:STATUS?
-> :INPUT:EQ:STATUS OFF
Description When the function is PulseWidth, AGC is OFF,
and EQ:STATus is turned OFF, the trigger
mode switches to manual trigger.
10.2 Commands
10
10-21

Advertisement

Table of Contents
loading

Table of Contents