YOKOGAWA TA220 User Manual page 23

Digital jitter meter
Table of Contents

Advertisement

2.3 Measurement Functions (Measurement Items)
D-to-C Jitter Excluding 2T <For the setup procedure, see section 4.1>
Example
Slope of the data signal: Both rising and falling
Slope of the clock signal: Rising
Data
signal
Clock
signal
2-6
This is one of the functions available for measuring D-to-C jitter. D-to-C jitter is
measured, excluding the phase difference of data signals having pulse widths of two
times the clock signal period (2T) or less. When the pulse width of the data signal is 2T
or less, the phase difference before and after that data signal is not measured.
The method for determining the jitter σ, jitter ratio σ/T, and average value is the same as
that for normal D-to-C jitter.
D-to-C jitter excluding 2T cannot be measured by products with suffix code -BDS.
Not measured
Not measured
2T
3T
Measured
Not measured
3T
2T
Measured
Not measured
3T
Measured
3T
4T
IM 704610-01E